Title: Evaluation of the mechanical behaviour of a bipolar hip prosthesis under transient loading

Authors: Rabiteja Patra; Harish Chandra Das; Shreeshan Jena

Addresses: Department of Mechanical Engineering, Siksha 'O' Anusandhan University, Bhubaneswar, Odisha, India ' Department of Mechanical Engineering, National Institute of Technology Meghalaya, Shillong, India ' Department of Biotechnology and Medical Engineering, National Institute of Technology Rourkela, Odisha, India

Abstract: Most of the studies available in the open literature make use of static analysis and discretisation of the load components for studying the mechanical behaviour of implants and prosthesis. The present study discusses the effect of time-varying loading on the prosthesis and femur bone assembly. The solid model of the femur bone was reconstructed using femur bone slices obtained from computed tomography (CT). The components of the hip joint forces and moments were applied at the femoral head of the prosthesis. The results from the present study were compared with the data from literature, and the present study shows that a time-varying loading analysis can provide much more realistic information about the prosthesis as compared to the prevailing use of static analyses.

Keywords: transient loading; CT; gait; finite element analysis; femoral prosthesis.

DOI: 10.1504/IJBET.2022.124190

International Journal of Biomedical Engineering and Technology, 2022 Vol.39 No.3, pp.314 - 326

Received: 26 Feb 2019
Accepted: 05 Jul 2019

Published online: 18 Jul 2022 *

Full-text access for editors Full-text access for subscribers Purchase this article Comment on this article